ΔΔCt vs the Pfaffl method vs the standard-curve method
The three methods differ in how they treat amplification efficiency. ΔΔCt fixes it at 100% (doubling every cycle), Pfaffl corrects with an efficiency measured for each gene, and the standard-curve method reads quantities straight from a dilution series. Computing the same data three ways shows how much it matters.
The three methods at a glance
| ΔΔCt (Livak) | Pfaffl | Standard curve | |
|---|---|---|---|
| Efficiency | Assumed 100% (E = 2) | Measured E for each gene | Built into the curve |
| You need | Target and reference Ct | Ct plus each gene’s efficiency | A dilution series on the plate |
| Output | Ratio to control | Ratio to control | Quantity per sample, and ratio |
| Best when | Efficiencies 90–110% and similar | Efficiencies differ or are far from 100% | Efficiency varies by plate, or quantities are needed |
- Efficiency from the slope: 10−1/slope − 1 (amplification factor E = 1 + efficiency). A slope of −3.32 is 100%
- Typical acceptance: 90–110% efficiency and R² ≥ 0.98 (BenchFig flags values outside these)
- As a rule of thumb, if the two genes differ by more than ~5 points, use Pfaffl or a standard curve instead of ΔΔCt
- Whatever the method, test on a log scale (ΔCt or log₂ ratio)
Calculating amplification efficiency
Run a 10-fold dilution series, plot Ct against log10(starting quantity) and fit a straight line. At 100% efficiency, 10 times more template comes up log210 ≈ 3.32 cycles earlier, so the slope is −3.32.
Here IL6 has 95.2% efficiency (factor 1.952) and GAPDH 99.7% (1.997). Both are within 90–110%, but they differ by about 4.5 points. For building the curve, quantifying unknowns and a free Excel template, see the qPCR standard curve guide.
The same data, three ways
Mean Ct values: control IL6 30.10 and GAPDH 18.40; treated IL6 25.60 and GAPDH 18.50. How many times higher is IL6 in the treated group?
ΔΔCt
Pfaffl
For target and reference separately, take the Ct difference “control − treated” and raise each gene’s own amplification factor to that power.
Standard curve
Convert each Ct to a quantity with its curve (quantity = 10(Ct − intercept)/slope), normalize target to reference, then divide treated by control. The result is 21.73-fold, identical to Pfaffl with the same efficiencies.
| Method | Treated / control | Difference from ΔΔCt |
|---|---|---|
| ΔΔCt (E = 2 assumed) | 24.25 | — |
| Pfaffl (measured E) | 21.73 | -10.4% |
| Standard curve | 21.73 | -10.4% |
Efficiency differences compound with the Ct difference. Even at ΔCt = 4.5 cycles, 95% versus 100% efficiency gives 1.9524.5 ≈ 20.28 versus 24.5 ≈ 22.63, about 10% apart.
Which method to choose
- Efficiencies within 90–110% and close to each other: ΔΔCt is fine. Report that you checked efficiency, and the values (MIQE guidelines).
- Efficiencies differ between genes or are far from 100%: Pfaffl. Take efficiency from a standard curve or from a separate experiment. In BenchFig, choosing “Correct for efficiency (Pfaffl)” uses the efficiency from the curves in the same file, and lets you type an efficiency (%) for genes without standards.
- Efficiency varies by plate, or you want the quantities themselves: the standard-curve method. It costs wells for a dilution series on each plate but does not rely on an efficiency assumption.
With any method, test on the log scale rather than on the ratio itself (see “Which values to test” in How to calculate ΔΔCt). For three or more groups, see choosing a statistical test.
FAQ
Why is my efficiency above 100%?
Usually inhibition in the most concentrated dilutions (delayed Ct flattens the slope), pipetting error, or non-specific products such as primer dimers. Check whether dropping the most concentrated points brings it into 90–110%, and look for extra peaks in the melt curve.
Do I need a standard curve on every plate?
For the standard-curve method, yes: standards should be on the same plate. For Pfaffl, an efficiency measured once with the same primers and conditions is often reused. Either way, report the efficiency and how it was obtained.
Do Pfaffl and the standard-curve method give the same result?
If the efficiencies come from the same curves, they are mathematically identical (21.73 in the example below). The standard-curve method also gives a quantity for each sample; Pfaffl only needs the efficiency, not standards on every plate.
How does Pfaffl handle two reference genes?
Compute ErefΔCt for each reference gene and divide by their geometric mean (the qBase approach of Hellemans et al.). BenchFig does this, and in the standard-curve method normalizes to the geometric mean of the reference quantities.
Related guides
Product and journal names are trademarks or registered trademarks of their respective owners.